Introduction Agilent N9020A, the MXA Option 513 provides frequency coverage up to 13.6 GHz. Parameters Frequency range Frequency reference Accuracy: ± [(time since last adjustment x aging rate) + temperature stability + calibration accuracy] Frequency readout accuracy (start, stop, center, marker) ± (marker frequency x frequency reference accuracy + 0.25 % x span + 5 % x RBW + 2 Hz + 0.5 x horizontal resolution 1) Marker frequency counter -
Accuracy: ± (marker frequency x frequency reference accuracy + 0.100 Hz) -
Delta counter accuracy: ± (delta frequency x frequency reference accuracy + 0.141 Hz) -
Counter resolution: 0.001 Hz Frequency span (FFT and swept mode) -
Range: 0 Hz (zero span), 10 Hz to maximum frequency of instrument -
Resolution: 2 Hz -
Accuracy: Swept ± (0.25 % x span + horizontal resolution ) -
FFT ± (0.10 % x span + horizontal resolution) Resolution bandwidth (RBW) Range: (–3.01 dB bandwidth) 1 Hz to 3 MHz (10 % steps), 4, 5, 6, 8 MHz
|